CVE-2026-7251

CRITICAL EPSS 38.9%
Published May 26, 20261mo ago · Modified Jun 17, 20262w ago
9.3 CVSS 4.0
Critical
Find Similar
Published May 26, 2026 1mo ago
Last Modified Jun 17, 2026 2w ago

Description

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

CVSS Details

Base Score
9.3
Exploitability
Impact
Vector string
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Attack Vector Network
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X

Threat Intelligence

EPSS Exploit Probability
38.9% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-259

References 3

  • github.com https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-146-01.json
  • cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-146-01
  • eppendorf.com https://www.eppendorf.com/software-downloads

Remediation

No remediation data recorded yet

Check vendor advisories and the NVD entry for patch availability.