Description
Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
CVSS Details
Base Score
Exploitability
Impact
Vector string
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X Attack Vector Network
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X
Threat Intelligence
EPSS Exploit Probability
38.9% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available
Weaknesses 1
CWE-259
References 3
- github.com https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-146-01.json
- cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-146-01
- eppendorf.com https://www.eppendorf.com/software-downloads
Remediation
No remediation data recorded yet
Check vendor advisories and the NVD entry for patch availability.