CVE-2026-0611

CRITICAL EPSS 47.1%
Published Jun 2, 20264w ago · Modified Jun 17, 20262w ago
9.2 CVSS 4.0
Critical
Find Similar
Published Jun 2, 2026 4w ago
Last Modified Jun 17, 2026 2w ago

Description

Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.

CVSS Details

Base Score
9.2
Exploitability
Impact
Vector string
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Attack Vector Network
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X

Threat Intelligence

EPSS Exploit Probability
47.1% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-306 Missing Authentication for Critical Function Authentication

References 3

  • spacelabshealthcare.com https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel
  • spacelabshealthcare.com https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf
  • vulncheck.com https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting

Remediation

No remediation data recorded yet

Check vendor advisories and the NVD entry for patch availability.