CVE-2025-4395

MEDIUM EPSS 18.6%
Published Jul 24, 202511mo ago · Modified Jun 17, 20262w ago
6.8 CVSS 3.1
Medium
Find Similar
Published Jul 24, 2025 11mo ago
Last Modified Jun 17, 2026 2w ago

Description

Medtronic MyCareLink Patient Monitor has a built-in user account with an empty password, which allows an attacker with physical access to log in with no password and access modify system functionality. This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025

CVSS Details

Base Score
6.8
Exploitability
0.9
Impact
5.9
Vector string
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack Vector Physical
Attack Complexity Low
Privileges Required None
User Interaction None
Scope Unchanged
Confidentiality High
Integrity High
Availability High

Threat Intelligence

EPSS Exploit Probability
18.6% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-258

References 2

  • cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-205-01
  • medtronic.com https://www.medtronic.com/en-us/e/product-security/security-bulletins/mycarelink-patient-monitor-vulnerabilities.html

Remediation

No remediation data recorded yet

Check vendor advisories and the NVD entry for patch availability.