CVE-2025-15653
HIGH EPSS 6.6%
Published Jun 2, 20264w ago · Modified Jun 17, 20262w ago
7.0 CVSS 4.0
Published Jun 2, 2026 4w ago
Last Modified Jun 17, 2026 2w ago
Description
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
CVSS Details
Base Score
Exploitability
Impact
Vector string
CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X Attack Vector Physical
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X
Threat Intelligence
EPSS Exploit Probability
6.6% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available
Weaknesses 1
CWE-668
References 2
- static.draeger.com https://static.draeger.com/security/download/Product-Security-Advisory-25-349-Zeus.pdf
- vulncheck.com https://www.vulncheck.com/advisories/dr-ger-zeus-ie-anesthesia-workstation-usb-interface-privilege-escalation
Remediation
No remediation data recorded yet