Description

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.  Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.                                                                                                                                                                                        Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

CVSS Details

Base Score
6.8
Exploitability
0.9
Impact
5.9
Vector string
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack Vector Physical
Attack Complexity Low
Privileges Required None
User Interaction None
Scope Unchanged
Confidentiality High
Integrity High
Availability High

Threat Intelligence

EPSS Exploit Probability
4.0% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 2

CWE-284
CWE-693

References 1

Remediation

No remediation data recorded yet