Description

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

CVSS Details

Base Score
8.2
Exploitability
Impact
Vector string
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Attack Vector Network
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X

Threat Intelligence

EPSS Exploit Probability
50.8% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-359

References 4

  • bleepingcomputer.com https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/
  • cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
  • cisa.gov https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor
  • fda.gov https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication

Remediation

No remediation data recorded yet

Check vendor advisories and the NVD entry for patch availability.