CVE-2025-0626
HIGH EPSS 62.0%
Published Jan 30, 20251y ago · Modified Jun 17, 20262w ago
7.7 CVSS 4.0
Published Jan 30, 2025 1y ago
Last Modified Jun 17, 2026 2w ago
Description
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
CVSS Details
Base Score
Exploitability
Impact
Vector string
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X Attack Vector Network
Attack Complexity Low
Privileges Required None
User Interaction P
Scope X
Threat Intelligence
EPSS Exploit Probability
62.0% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available
Weaknesses 1
CWE-912
References 4
- bleepingcomputer.com https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/
- cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
- cisa.gov https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor
- fda.gov https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication
Remediation
No remediation data recorded yet
Check vendor advisories and the NVD entry for patch availability.