CVE-2025-0626

HIGH EPSS 62.0%
Published Jan 30, 20251y ago · Modified Jun 17, 20262w ago
7.7 CVSS 4.0
High
Find Similar
Published Jan 30, 2025 1y ago
Last Modified Jun 17, 2026 2w ago

Description

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

CVSS Details

Base Score
7.7
Exploitability
Impact
Vector string
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Attack Vector Network
Attack Complexity Low
Privileges Required None
User Interaction P
Scope X

Threat Intelligence

EPSS Exploit Probability
62.0% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-912

References 4

  • bleepingcomputer.com https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/
  • cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
  • cisa.gov https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor
  • fda.gov https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication

Remediation

No remediation data recorded yet

Check vendor advisories and the NVD entry for patch availability.