CVE-2024-9834
CRITICAL EPSS 3.5%
Published Nov 14, 20241y ago · Modified Jun 17, 20262w ago
9.3 CVSS 3.1
Published Nov 14, 2024 1y ago
Last Modified Jun 17, 2026 2w ago
Description
Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
CVSS Details
Base Score
Exploitability
Impact
Vector string
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H Attack Vector Local
Attack Complexity Low
Privileges Required None
User Interaction None
Scope Changed
Confidentiality High
Integrity High
Availability High
Threat Intelligence
EPSS Exploit Probability
3.5% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available
Weaknesses 1
CWE-319
References 1
- cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01
Remediation
No remediation data recorded yet
Check vendor advisories and the NVD entry for patch availability.