CVE-2024-9832
CRITICAL EPSS 7.5%
Published Nov 14, 20241y ago · Modified Jun 17, 20262w ago
9.3 CVSS 3.1
Published Nov 14, 2024 1y ago
Last Modified Jun 17, 2026 2w ago
Description
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
CVSS Details
Base Score
Exploitability
Impact
Vector string
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H Attack Vector Local
Attack Complexity Low
Privileges Required None
User Interaction None
Scope Changed
Confidentiality High
Integrity High
Availability High
Threat Intelligence
EPSS Exploit Probability
7.5% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available
Weaknesses 1
CWE-307
References 1
- cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01
Remediation
No remediation data recorded yet
Check vendor advisories and the NVD entry for patch availability.