CVE-2024-48974
CRITICAL EPSS 5.0%
Published Nov 14, 20241y ago · Modified Jun 17, 20262w ago
9.3 CVSS 3.1
Published Nov 14, 2024 1y ago
Last Modified Jun 17, 2026 2w ago
Description
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
CVSS Details
Base Score
Exploitability
Impact
Vector string
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H Attack Vector Local
Attack Complexity Low
Privileges Required None
User Interaction None
Scope Changed
Confidentiality High
Integrity High
Availability High
Threat Intelligence
EPSS Exploit Probability
5.0% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available
Weaknesses 1
CWE-494
References 1
- cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01
Remediation
No remediation data recorded yet
Check vendor advisories and the NVD entry for patch availability.