CVE-2024-48973
CRITICAL EPSS 12.6%
Published Nov 14, 20241y ago · Modified Jun 17, 20262w ago
9.3 CVSS 3.1
Published Nov 14, 2024 1y ago
Last Modified Jun 17, 2026 2w ago
Description
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
CVSS Details
Base Score
Exploitability
Impact
Vector string
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H Attack Vector Local
Attack Complexity Low
Privileges Required None
User Interaction None
Scope Changed
Confidentiality High
Integrity High
Availability High
Threat Intelligence
EPSS Exploit Probability
12.6% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available
Weaknesses 1
CWE-1263
References 1
- cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01
Remediation
No remediation data recorded yet