CVE-2024-48970

CRITICAL EPSS 11.6%
Published Nov 14, 20241y ago · Modified Jun 17, 20261w ago
9.3 CVSS 3.1
Critical
Find Similar
Published Nov 14, 2024 1y ago
Last Modified Jun 17, 2026 1w ago

Description

The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure.

CVSS Details

Base Score
9.3
Exploitability
2.5
Impact
6.0
Vector string
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Attack Vector Local
Attack Complexity Low
Privileges Required None
User Interaction None
Scope Changed
Confidentiality High
Integrity High
Availability High

Threat Intelligence

EPSS Exploit Probability
11.6% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-1191

References 1

  • cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01

Remediation

No remediation data recorded yet

Check vendor advisories and the NVD entry for patch availability.