CVE-2024-48967

CRITICAL EPSS 44.8%
Published Nov 14, 20241y ago · Modified Jun 17, 20262w ago
10.0 CVSS 3.1
Critical
Find Similar
Published Nov 14, 2024 1y ago
Last Modified Jun 17, 2026 2w ago

Description

The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.

CVSS Details

Base Score
10.0
Exploitability
3.9
Impact
6.0
Vector string
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Attack Vector Network
Attack Complexity Low
Privileges Required None
User Interaction None
Scope Changed
Confidentiality High
Integrity High
Availability High

Threat Intelligence

EPSS Exploit Probability
44.8% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-778

References 1

  • cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01

Remediation

No remediation data recorded yet

Check vendor advisories and the NVD entry for patch availability.