CVE-2024-48966

CRITICAL EPSS 47.6%
Published Nov 14, 20241y ago · Modified Jun 17, 20262w ago
10.0 CVSS 3.1
Critical
Find Similar
Published Nov 14, 2024 1y ago
Last Modified Jun 17, 2026 2w ago

Description

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

CVSS Details

Base Score
10.0
Exploitability
3.9
Impact
6.0
Vector string
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Attack Vector Network
Attack Complexity Low
Privileges Required None
User Interaction None
Scope Changed
Confidentiality High
Integrity High
Availability High

Threat Intelligence

EPSS Exploit Probability
47.6% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-306 Missing Authentication for Critical Function Authentication

References 1

  • cisa.gov https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01

Remediation

No remediation data recorded yet