CVE-2022-4992

HIGH EPSS 5.6%
Published Jun 2, 20261mo ago · Modified Jun 17, 20262w ago
8.8 CVSS 4.0
High
Find Similar
Published Jun 2, 2026 1mo ago
Last Modified Jun 17, 2026 2w ago

Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

CVSS Details

Base Score
8.8
Exploitability
Impact
Vector string
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Attack Vector Network
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X

Threat Intelligence

EPSS Exploit Probability
5.6% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-345

References 2

  • static.draeger.com https://static.draeger.com/security/download/PSA-19-255-03_Product-Security-Advisory-IACS.pdf
  • vulncheck.com https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofed-network-message-handling-dos-tampering

Remediation

No remediation data recorded yet

Check vendor advisories and the NVD entry for patch availability.