CVE-2019-25724
HIGH EPSS 7.2%
Published Jun 2, 20261mo ago · Modified Jun 17, 20262w ago
7.1 CVSS 4.0
Published Jun 2, 2026 1mo ago
Last Modified Jun 17, 2026 2w ago
Description
Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.
CVSS Details
Base Score
Exploitability
Impact
Vector string
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X Attack Vector Adjacent
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X
Threat Intelligence
EPSS Exploit Probability
7.2% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available
Weaknesses 1
CWE-400 Uncontrolled Resource Consumption Resource Mgmt
References 2
- static.draeger.com https://static.draeger.com/security/download/2019-277-02-M300-VG2x-Security-Advisory.pdf
- vulncheck.com https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-denial-of-service
Remediation
No remediation data recorded yet
Check vendor advisories and the NVD entry for patch availability.