CVE-2019-25724

HIGH EPSS 7.2%
Published Jun 2, 20261mo ago · Modified Jun 17, 20262w ago
7.1 CVSS 4.0
High
Find Similar
Published Jun 2, 2026 1mo ago
Last Modified Jun 17, 2026 2w ago

Description

Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

CVSS Details

Base Score
7.1
Exploitability
Impact
Vector string
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Attack Vector Adjacent
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X

Threat Intelligence

EPSS Exploit Probability
7.2% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-400 Uncontrolled Resource Consumption Resource Mgmt

References 2

  • static.draeger.com https://static.draeger.com/security/download/2019-277-02-M300-VG2x-Security-Advisory.pdf
  • vulncheck.com https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-denial-of-service

Remediation

No remediation data recorded yet

Check vendor advisories and the NVD entry for patch availability.