CVE-2019-25723

MEDIUM EPSS 14.5%
Published Jun 2, 20261mo ago · Modified Jun 17, 20262w ago
6.3 CVSS 4.0
Medium
Find Similar
Published Jun 2, 2026 1mo ago
Last Modified Jun 17, 2026 2w ago

Description

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

CVSS Details

Base Score
6.3
Exploitability
Impact
Vector string
CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Attack Vector Network
Attack Complexity High
Privileges Required None
User Interaction None
Scope X

Threat Intelligence

EPSS Exploit Probability
14.5% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-1286

References 2

  • static.draeger.com https://static.draeger.com/security/download/2020-06-PSA-20-169-01-Security-Advisory-Perseus-A500-SW20n.pdf
  • vulncheck.com https://www.vulncheck.com/advisories/dr-ger-perseus-a500-dos-via-medibus-interface

Remediation

No remediation data recorded yet

Check vendor advisories and the NVD entry for patch availability.