Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.

CVSS Details

Base Score
7.1
Exploitability
Impact
Vector string
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Attack Vector Adjacent
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X

Threat Intelligence

EPSS Exploit Probability
8.5% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-400 Uncontrolled Resource Consumption Resource Mgmt

References 2

  • static.draeger.com https://static.draeger.com/security/download/2019-09-13-Cybersecurity-M300-VGxx-Security-advisory-update_v1-3.pdf
  • vulncheck.com https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service

Remediation

No remediation data recorded yet

Check vendor advisories and the NVD entry for patch availability.