CVE-2019-25721
HIGH EPSS 8.5%
Published Jun 2, 20263w ago · Modified Jun 17, 20261w ago
7.1 CVSS 4.0
Published Jun 2, 2026 3w ago
Last Modified Jun 17, 2026 1w ago
Description
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
CVSS Details
Base Score
Exploitability
Impact
Vector string
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X Attack Vector Adjacent
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X
Threat Intelligence
EPSS Exploit Probability
8.5% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available
Weaknesses 1
CWE-400 Uncontrolled Resource Consumption Resource Mgmt
References 2
- static.draeger.com https://static.draeger.com/security/download/2019-09-13-Cybersecurity-M300-VGxx-Security-advisory-update_v1-3.pdf
- vulncheck.com https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service
Remediation
No remediation data recorded yet
Check vendor advisories and the NVD entry for patch availability.