CVE-2019-25719
HIGH EPSS 3.1%
Published Jun 2, 20261mo ago · Modified Jun 17, 20262w ago
8.8 CVSS 4.0
Published Jun 2, 2026 1mo ago
Last Modified Jun 17, 2026 2w ago
Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
CVSS Details
Base Score
Exploitability
Impact
Vector string
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X Attack Vector Network
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X
Threat Intelligence
EPSS Exploit Probability
3.1% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available
Weaknesses 1
CWE-924
References 2
- static.draeger.com https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf
- vulncheck.com https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling
Remediation
No remediation data recorded yet
Check vendor advisories and the NVD entry for patch availability.