CVE-2019-25718
HIGH EPSS 2.3%
Published Jun 1, 20264w ago · Modified Jun 17, 20261w ago
8.6 CVSS 4.0
Published Jun 1, 2026 4w ago
Last Modified Jun 17, 2026 1w ago
Description
Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.
CVSS Details
Base Score
Exploitability
Impact
Vector string
CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X Attack Vector Local
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X
Threat Intelligence
EPSS Exploit Probability
2.3% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available
Weaknesses 1
CWE-451
References 2
- static.draeger.com https://static.draeger.com/security/download/2019-01-22-draeger-infinity-delta-vf10-1-security-advisory.pdf
- vulncheck.com https://www.vulncheck.com/advisories/dr-ger-infinity-explorer-c700-privilege-escalation-via-kiosk-mode-bypass
Remediation
No remediation data recorded yet
Check vendor advisories and the NVD entry for patch availability.