Description

Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.

CVSS Details

Base Score
7.1
Exploitability
Impact
Vector string
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Attack Vector Adjacent
Attack Complexity Low
Privileges Required None
User Interaction None
Scope X

Threat Intelligence

EPSS Exploit Probability
10.5% percentile
Exploit & Patch Status
No Known Exploit
No Patch Available

Weaknesses 1

CWE-15

References 2

  • static.draeger.com https://static.draeger.com/security/download/2019-01-22-draeger-infinity-delta-vf10-1-security-advisory.pdf
  • vulncheck.com https://www.vulncheck.com/advisories/dr-ger-infinity-delta-kappa-patient-monitor-dos-via-malformed-network-packet

Remediation

No remediation data recorded yet

Check vendor advisories and the NVD entry for patch availability.